Model | GS-U50V | GS-U75V | GS-U100V |
Chamber Size | |||
Volume | 50L | 75L | 100L |
Chamber Size | φ340×550mm | φ400×600mm | φ440×650mm |
Pressure Performance | |||
Working Pressure | 0.22Mpa | ||
Max. Working Pressure | 0.23Mpa | ||
Temperature & Timming Performance | |||
Working Temperature | 134℃ | ||
Temperature Adjust Range | 0-134℃ | ||
Heat Average | ≤±1℃ | ||
Timer Range | 0-99min | ||
Display and Controller | |||
Display | LCD display of working status, touch-type keys. | ||
Controller | Micro-computer processor controller | ||
Sensor | PT100 | ||
Electric Supply | |||
Power Consumption | 3KW | 4.5KW | 4.5KW |
Power Supply | AC220V 50HZ/60HZ | ||
Material and Construction | |||
External Material | 304 stainless steel | ||
Internal Material | 304 stainless steel | ||
Basket Material and Qty | 304 stainless steel baskets, 2pcs | ||
Size | |||
Overall Size(mm) | 690×470×1140 | 730×510×1270 | 750×550×1305 |
Transport Size(mm) | 760×550×1290 | 820×600×1380 | 840×640×1410 |
N.W/G.W | 100/120Kg | 118/140Kg | 128/160Kg |
Application | |||
The vertical pressure steam sterilizers are assembled with heating system, microcomputer controlled system, over heat and over pressure protection system, which are reliable for the sterilizing effects, convenient of operation and energy conservation. They are ideal equipments for clinics, scientific research institutes and other organizations to sterilizer surgical instruments, fabrics, glasses, culture media etc. | |||
Characteristics | |||
• Fully automatic process of water inlet, pulse vacuum, heating. sterilizing, draining. exhausting, vacuum drying, and auto shut off with beep alarm after sterilization. • 3 times pulse vacuum (the vacuum rate would reach -0.081mpa) ensures the complete extracting of inner air from the dressing and A, B type hollow appliance. • LCD display with microcomputer controller, dynamically shows the temperature, pressure time ,vacuum rate and the cycle status. • Steam-water inner circulation system: no steam discharge and the environment for the sterilizing will be clean and dry. • 4 fixed programs for the appliance, medical dressing, rubber and fluid; and settable customized program. • Multiple pulse vacuum drying after sterilizing, which ensures the good sterilizing result. Accessories • 0.22pum bacteria filter ensures the air enterring the vacuumed chamber is sterilized, avoiding the second time contamination. • Equipped with the PY/ TT testing connector, easy for running the testing. • Adopted with high-heating-efficiency immersed heating tube. • Hand wheel type for quick-open door with selfinflating seal. |